Skip to main content

Table 3 Circuit lifetime and CRRT interruption causes

From: Continuous venovenous hemodiafiltration with a low citrate dose regional anticoagulation protocol and a phosphate-containing solution: effects on acid–base status and phosphate supplementation needs

 

Protocol A (n = 102)

Protocol B (n = 138)

Overall (n = 240)

CIRCUIT LIFETIME

   

Mean ± SD, h

51.8 ± 36.5

53 ± 32.6

52.5 ± 34.2

Median (IQR), h

44.5 (24.0–72.0)

47.5 (24.0–78.5)

47.5 (24.0–75.0)

> 24 h

77%

76%

76.7%

> 48 h

50%

50%

50%

> 72 h

28%

33%

31.2%

CRRT STOPPING CAUSES

   

CVC malfunction, n (%)

40 (39.2%)

49 (35.5%)

89 (37.1%)

Alarm handling/ technical issues, n (%)

26 (25.4%)

31 (22.5%)

57 (23.8%)

Scheduled, n (%)

24 (23.6%)

36 (26.1%)

60 (25%)

Medical procedures, n (%)

12 (11.8%)

13 (9.4%)

25 (10.4%)

Clotting, n (%)

0 (0%)

9 (6.5%)

9 (3.7%)

  1. Data are expressed as mean ± SD or median (IQR) or percentage.