Skip to main content

Table 2 The concentration of vancomycin and gentamicin at different time interval in different conditions

From: Determination of vancomycin and gentamicin clearance in an in vitro, closed loop dialysis system

Condition

Filter

Sol

Time (min)

BFR (mL/min)

DFR (mL/min)

Vanco (ug/mL)

TG (ug/mL)

C1

Poly

SBS

0

400

600

20.04 ± 0.86

26.65 ± 1.69

Poly

SBS

180

400

600

BLQ

0.51 ± 0

Poly

Dial

0

400

600

BLQ

BLQ

Poly

Dial

180

400

600

1.35 ± 0.41

1.31 ± 0.18

C2

Poly

SBS

0

500

800

31.91 ± 2.5

35.93 ± 0.13

Poly

SBS

180

500

800

BLQ

0.53 ± 0.02

Poly

Dial

0

500

800

BLQ

BLQ

Poly

Dial

180

500

800

1.06 ± 0.05

0.99 ± 0

C3

Opt*

SBS

0

400*

600*

23.74 ± 5.27

25.41 ± 3.47

Opt

SBS

180

400

600

BLQ

0.52 ± 0.01

Opt

Dial

0

400

600

BLQ

BLQ

Opt

Dial

180

400

600

1.20 ± 1

1.14 ± 0.93

C4

Opt

SBS

0

500

800

30.75 ± 0.43

36.97 ± 0.18

Opt

SBS

180

500

800

BLQ

0.51 ± 0

Opt

Dial

0

500

800

BLQ

BLQ

Opt

Dial

180

500

800

0.91 ± 0.08

0.94 ± 0.15

  1. UFR is constant at rate of 300 ml/hr except for the one condition. *UFR was 250 mL/hr. Sol: Solution; Poly: Polyflux filter; Opt: Optiflux filter; Dial: Dialysate; Vanco: Mean vancomycin concentration; TG: Mean total gentamicin concentration; BLQ: Below the lower limit of quantification.