From: Determination of vancomycin and gentamicin clearance in an in vitro, closed loop dialysis system
Condition | Filter | Sol | BFR (mL/min) | DFR (mL/min) | Vanco (mg) | Gent (mg) | ClVanc(mL/min) | ClGent(mL/min) |
---|---|---|---|---|---|---|---|---|
C1 | Poly | Dial | 400 | 600 | 147.0 ± 44.7 | 142.7 ± 20 | 82.1 | 58.2 |
C2 | Poly | Dial | 500 | 800 | 152.9 ± 7.2 | 142.7 ± 1.02 | 53.3 | 43.5 |
C3 | Opt* | Dial | 400* | 600* | 131.0 ± 108.4* | 124.0 ± 101.5* | 65.3* | 56.6* |
C4 | Opt | Dial | 500 | 800 | 131.9 ± 12.3 | 137.0 ± 21.5 | 47.7 | 40.6 |