Skip to main content

Table 3 Summary of the quantity of vancomycin and gentamicin using the mass balance equation and dialytic clearance

From: Determination of vancomycin and gentamicin clearance in an in vitro, closed loop dialysis system

Condition

Filter

Sol

BFR (mL/min)

DFR (mL/min)

Vanco (mg)

Gent (mg)

ClVanc(mL/min)

ClGent(mL/min)

C1

Poly

Dial

400

600

147.0 ± 44.7

142.7 ± 20

82.1

58.2

C2

Poly

Dial

500

800

152.9 ± 7.2

142.7 ± 1.02

53.3

43.5

C3

Opt*

Dial

400*

600*

131.0 ± 108.4*

124.0 ± 101.5*

65.3*

56.6*

C4

Opt

Dial

500

800

131.9 ± 12.3

137.0 ± 21.5

47.7

40.6

  1. UFR is constant at rate of 300 ml/hr except for the one condition. *UFR was 250 ml/hr. Sol: Solution; Poly: Polyflux filter; Opt: Optiflux filter; Dial: Dialysate; Vanco: Vancomycin concentration; Gent: Total Gentamicin concentration; BLQ: Below the lower limit of quantification; ClVanc: Mean Vancomycin Clearance; ClGent: Mean Gentamicin clearance.