Table 1 Iron sucrose removal by high-flux or high-efficiency hemodialysis membranes over 240 minutes. (System A)

From: Determination of iron sucrose (Venofer) or iron dextran (DexFerrum) removal by hemodialysis: an in-vitrostudy

  Iron Sucrose (Venofer®)
Study condition SBS % iron concentration change
(mean ± SD)
Khd (hr-1)
(mean ± SD)
Dialysate Iron Recovery (mg)
(mean ± SD)
Clearance (ml/min)
(mean ± SD)
F80: 0 ml/hr UFR 13.55 ± 1.56 0.0006 ± 0.0001 0.00 ± 0.00 3.62 ± 0.45
F80: 0 ml/hr UFR Re-run* 6.99 ± 8.55 0.0003 ± 0.0004 0.00 ± 0.00 1.85 ± 2.29
F80: 500 ml/hr UFR -16.03 ± 8.57 -0.0006 ± 0.0003 0.00 ± 0.00 -2.46 ± 1.23
CA210: 0 ml/hr UFR -6.20 ± 1.78 -0.0003 ± 0.0001 0.00 ± 0.00 -1.49 ± 0.42
CA210: 500 ml/hr UFR -14.79 ± 31.51 -0.0003 ± 0.0019 0.00 ± 0.00 -6.87 ± 7.74
  1. SBS = simulated blood solution; UFR = ultrafiltration rate; Khd = hemodialysis elimination rate; % = percent; * = The re-run study condition was an unique experiment that replicated all conditions present in the original F80: 0 ml/hr UFR experiment. † = Clearance values that are positive relate to accumulation of substance from the system; negative clearance values represent removal of substance from the system.