From: A comparison of temporal artery thermometers with internal blood monitors to measure body temperature during hemodialysis

Fig. 2

Bland-Altman Plot showing level of agreement between the temporal artery (TA) and dialysis machine (DM) thermometers. Of the 94 data points, 3 (3.2%) are outside the limits of agreement (− 0.57 and 1.12 °C). The mean difference (dotted line) is 0.27 °C and is consistent over the average temperatures obtained in this study. Temperatures measured using the dialysis machine were rounded to single digit. Larger points represent multiple patients with identical values