Table 2 Body temperature of hemodialysis patients measured by the temporal artery thermometer and the internal blood monitor of the dialysis machine for each dataset

From: A comparison of temporal artery thermometers with internal blood monitors to measure body temperature during hemodialysis

Dataseta

Thermometerb

Mean BT (°C)

95% CI

p-valuec

A

Temporal Artery

36.7

36.6–36.8

< 0.001

Dialysis Machine

36.42

36.33–36.51

Mean Difference

0.27

0.18–0.37

 

B

Temporal Artery

36.7

36.6–36.8

< 0.001

Dialysis Machine

36.42

36.33–36.50

Mean Difference

0.27

0.18–0.37

 

C

Temporal Artery

36.7

36.6–36.7

< 0.001

Dialysis Machine

36.49

36.41–36.56

Mean Difference

0.21

0.13–0.28

 
  1. BT body temperature
  2. aA: first measurement only, irrespective of time (n = 94); B: first measurement only, within first 15 min of starting treatment (n = 87); C: all measurements (maximum of 3 per thermometer type), within first 15 min of starting treatment (n = 87)
  3. bDifference = (TA – DM)
  4. cPaired t-test comparing the mean temporal artery and dialysis machine temperatures