Table 4 Comparisons of clinical outcomes among the study groups after connecting additional pressure control lines

From: Using additional pressure control lines when connecting a continuous renal replacement therapy device to an extracorporeal membrane oxygenation circuit

Clinical outcomes

No line group

(n = 16)

Single line group

(n = 36)

Double line group

(n = 118)

P value

Duration of ECMO support, hours

80 (53–467)

93 (32–189)

184 (93–501)

0.003

Successful weaning from ECMO

6 (37.5)

17 (47.2)

72 (61.0)

0.103

ECMO complication

 Gastrointestinal bleeding

2 (12.5)

1 (2.9)

16 (13.8)

0.201

 ECMO site bleeding

2 (12.5)

8 (22.9)

15 (12.9)

0.339

 Cerebral infarction

2 (12.5)

4 (11.4)

3 (2.6)

0.053

 Sepsis

3 (18.8)

5 (14.3)

10 (8.6)

0.356

In-hospital mortality

11 (68.8)

27 (75.0)

57 (48.3)

0.010†‡

  1. ECMO extracorporeal membrane oxygenation
  2. Superscripts †, ‡ indicate significant differences (p < 0.05) between the no line group and the double line group, and the single line group and the double line group, respectively