Skip to main content

Table 3 KDIGO Hepatitis C guideline summary of hygienic precautions for dialysis machines. Reproduced from reference [34]

From: Clinical practice guideline management of blood borne viruses within the haemodialysis unit

Hygienic precautions for dialysis machines

Definitions

The ‘transducer protector’ is a filter (normally a hydrophobic 0.2-mm filter) that is fitted between the pressure monitoring line of the extracorporeal circuit and the pressure monitoring port of the dialysis machine. The filter allows air to pass freely to the pressure transducer that gives the reading displayed by the machine, but it resists the passage of fluid. This protects the patient from microbiologic contamination (as the pressure monitoring system is not disinfected) and the machine from ingress of blood or dialysate. An external transducer protector is normally fitted to each pressure monitoring line in the blood circuit. A back-up filter is located inside the machine. Changing the internal filter is a technical job.

A ‘single-pass machine’ is a machine that pumps the dialysate through the dialyser and then to waste. In general, such machines do not allow fluid to flow between the drain pathway and the fresh pathway except during disinfection. ‘Recirculating’ machines produce batches of fluid that can be passed through the dialyser several times.

Transducer protectors

External transducer protectors should be fitted to the pressure lines of the extracorporeal circuit.

Before commencing dialysis, staff should ensure that the connection between the transducer protectors and the pressure-monitoring ports is tight as leaks can lead to wetting of the filter.

Transducer protectors should be replaced if the filter becomes wet, as the pressure reading may be affected. Using a syringe to clear the flooded line may damage the filter and increase the possibility of blood passing into the dialysis machine so it is essential to fit a new transducer protector to the monitoring line if this procedure has to be used.

If wetting of the filter occurs after the patient has been connected, the line should be inspected carefully to see if any blood has passed through the filter. If any fluid is visible on the machine side, the machine should be taken out of service at the end of the session so that the internal filter can be changed and the housing disinfected.

External cleaning

After each session, the exterior of the dialysis machine should be cleaned with a low-level disinfectant if not visibly contaminated.

If a blood spillage has occurred, the exterior should be disinfected with a commercially available tuberculocidal germicide or a solution containing at least 500 p.p.m. hypochlorite (a 1:100 dilution of 5% household bleach) if this is not detrimental to the surface of dialysis machines. Advice on suitable disinfectants, and the concentration and contact time required, should be provided by the manufacturer.

If blood or fluid is thought to have seeped into inaccessible parts of the dialysis machine (for example, between modules, behind blood pump), the machine should be taken out of service until it can be dismantled and disinfected.

Disinfection of the internal fluid pathways

It is not necessary for the internal pathways of a single-pass dialysis machines to be disinfected between patients, unless a blood leak has occurred, in which case both the internal fluid pathways and the dialysate-to-dialyser (Hansen) connectors should be disinfected before the next patient.

If machines are not subjected to an internal disinfection procedure, staff should ensure that sufficient time is available between patients for the external surfaces to be disinfected.

Machines with recirculating dialysate should always be put through an appropriate disinfection procedure between patients.