Skip to main content

Table 3 Clinically significant alarm data stratified by positive or negative COVID-19 status (n = number of treatments)

From: Extension of Tablo TrEatmeNt Duration (XTEND) study: successful 24 h prolonged therapy with Tablo in critical patients

Alarm Parameter

(Mean ± SD)

Total

Non-COVID-19

COVID-19

(n = 87)

(n = 74)

(n = 13)

Alarm Rate per Hour

0.5 ± 0.6

0.5 ± 0.6

0.7 ± 0.9

Alarm Resolution Time (seconds)

18 ± 18

16 ± 17

26 ± 20

Blood Pump Stoppage Time (minutes)

4.5 ± 5.2

3.7 ± 4.2

8.8 ± 7.7