Figure 3From: Exsanguination of a home hemodialysis patient as a result of misconnected blood-lines during the wash back procedure: A case report The saline bag found attached to the deceased’s dialysis machine. The weight of the bag was 3.3kg, implying the addition of 2.3L of the patient’s blood to the bag of saline during the wash back procedure.Back to article page