Table 2 Comparison of fistula and non-fistula measurements

From: Oscillometric pulse wave analysis for detecting low flow arteriovenous fistula

 

Overall Median

Fi versus nFi status

Parameter

MedianFi (n = 53)

min/max

MediannFi (n = 53)

min/max

p (Wilcoxon)

(Fi and nFi, n = 106)

Sensitivity

Specificity

p (Fisher's exact)

Slope2 [relative amplitude/ms]

-0,0015

-0.00408/ 0.00497

-0,00322

-0.00564/ -0.00183

 < 0.001

-0,00269

0,79

0,79

 < 0.001

Slope∑ [relative amplitude/ms]

0,01094

-0.00022/ 0.01383

0,01187

0.00893/ 0.01666

0,006

0,01146

0,64

0,64

0,003

AMP [relative volumetric change]

1850

473/ 5236

725

190/ 2180

 < 0.001

1030

0,81

0,81

 < 0.001

Vf [ml/min]

795

247/ 2128

58

2/ 168

 < 0.001

207

1

1

 < 0.001

Abrach. [mm2]

32

16/ 71

19

9/ 44

 < 0.001

25

0,7

0,72

 < 0.001

  1. Lower case index indicating fistula (Fi) and non-fistula (nFi) measurements, for discription of variables see method section of the main document