Table 3 Testing for low flow fistulas: ROC-Analysis

From: Oscillometric pulse wave analysis for detecting low flow arteriovenous fistula

Parameter

AUC

p

Lower Bound

Upper Bound

maximal Youden's-Index

corresponding cut-off

Sensitivity

Specificity

Positive predictive value

Negative predictive value

Slope2Fi

0,652

0,167

0,437

0,866

0,35

0,000026

0,8

0,56

0,27

0,93

∆Slope2

0,624

0,253

0,411

0,836

0,28

0,002744

0,73

0,44

0,21

0,89

Slope∑Fi

0,732

0,006

0,566

0,899

0,42

0,009639

0,86

0,56

0,29

0,95

∆Slope∑

0,662

0,163

0,434

0,889

0,32

0,000782

0,61

0,78

0,36

0,91

AMPFi

0,775

0,012

0,56

0,991

0,64

1138

0,86

0,78

0,44

0,96

∆AMP

0,823

0,002

0,622

1,024

0,71

416

0,93

0,78

0,46

0,98

  1. For discription of variables see method section of the main document, ROC receiver operator curve, presented by AUC and confidence interval bounds with p-value, n = 53: Values of measurements at the fistula arm (Slope2Fi, Slope∑Fi and AMPFi) as well as differences of measurements at the fistula and non-fistula arm (∆Slope2, ∆Slope∑ and ∆AMP) were analysed. A low flow fistula was defined by duplex measurements of flow < 500 ml/min at the fistula arm