Accuracy of iohexol plasma clearance for GFR-determination: a comparison between single and dual sampling

Abstract

Background

Current guidelines regarding plasma-sampling techniques for glomerular filtration rate (GFR) determination are inconsistent. Single-sample methods are commonly believed not to be precise enough to meet clinical demands. The present study compared the agreement between single- and dual- plasma sampling methods with a three-point plasma clearance of iohexol.

Methods

A total of 46 healthy volunteers and 124 chronic kidney disease (CKD) patients with varying degrees of renal dysfunction received 5 ml iohexol (300 mgI/ml) i.v. and plasma samples were drawn at 2-, 3- and 4-h post-injection. Plasma-iodine concentrations were detected by high-performance liquid chromatography (HPLC).

Results

Bias was similar among single-plasma sampling methods (SPSM) and dual-plasma sampling methods (DPSM). The best correlation was obtained from the 2- and 4-h DPSM (concordance correlation coefficient [CCC]: 0.9988) with none of the estimates differed by more than 30% from the reference GFR and only one (0.06%) estimate differed by more than 10% (P30, 100%; P10, 99.4%). SPSM using samples around 3- or 4-h demonstrated acceptable accuracy at a GFR level of ≥60 ml/min/1.73m2 (P30 = 100% and P10 > 75% for both measurements).

Conclusion

4-h SPSM is advantageous in clinical practice in subjects with GFR ≥ 60 ml/min/1.73m2. For patients with an expected GFR < 60 ml/min/1.73m2, a prolonged sampling time is more reliable.

Peer Review reports

Background

Calculated plasma clearance of iohexol after a single bolus injection correlates well with the “gold standard of renal function” inulin clearance and is recommended as a robust standard for evaluating renal function [1,2,3]. The mathematical model for the drug elimination curve is an open two-compartment system. The exogenous tracer for glomerular filtration rate (GFR) measurement is injected in the first compartment, equilibrates with the second compartment, and is excreted from the first compartment by glomerular filtration [4]. However, multiple blood samples beyond 6 h post-injection are needed to calculate the area under the time-concentration curve (AUC), if maximal precision and accuracy of the measurements are sought [5, 6], which would be rather time-consuming and therefore compromise recruitment and subject compliance. A growing need for simplification has led to the application of single-compartment models that need only two to three blood samples, which calculate AUC from the intercept and slope of the final slow disappearance curve. By the two-compartment correction introduced by Brochner-Mortensen [7], the slope-intercept method provides greater simplicity and sufficient accuracy to meet clinical demands.

Further simplified techniques requiring only one blood sample have also been developed [8,9,10,11,12,13,14]. Single-plasma sampling methods (SPSM) demonstrate acceptable accuracy, given that the single sample is drawn at a proper time point and there is knowledge about the distribution volume of the injected tracer. However, this method lacks the opportunity for quality control and has been reported inferior to dual-plasma sampling methods (DPSM) [9, 15,16,17,18].

In the present study, we compared the clearance values based on DPSM and SPSM with a three-point plasma clearance of iohexol (3 pt. iGFR), to specify whether 3 pt. iGFR measurement can be substituted by DPSM or SPSM, or which estimate is more adequate to be recommended in routine practice.

Methods

A total of 170 participants (89 females, 52.4%) including 46 healthy volunteers and 124 CKD patients with varying degrees of renal dysfunction were recruited in this study. Mean age was 43 years (range 21–87 years). The mean body-mass index and body-surface area were 24.2 ± 4.0 kg/m2 and 1.74 ± 0.2 m2 respectively. Mean serum-creatinine was 111.3 μmol/l (range 37.1–797.4 μmol/l). A large proportion of the studied patients were diagnosed with diabetes (14.1%) or hypertension (35.9%).

The participants were examined in a non-fasting state. Baseline blood sample was obtained at test day. A single bolus injection of 5 ml iohexol (Omnipaque, 300 mgI/ml, GE healthcare, Shanghai, China) was given at one side of the upper limb peripheral vein and then blood sample was drawn from the contralateral arm at 2 h, 3 h and 4 h after injection. Plasma-iodine concentrations were detected by high-performance liquid chromatography (Waters Alliance HPLC, Milford, USA).

The protocol was approved by the local ethics committee. All volunteers were informed and signed the consent form.

Dual- and multi-sample methods (Cl slope)

Based on the one-compartment model, plasma clearance of iohexol was calculated using the slope and the intercept of the regression equation in the final slow clearance.

$$ {\mathrm{C}\mathrm{l}}_{\mathrm{slope}}={\mathrm{Q}}_0/{\mathrm{C}}_1/\mathrm{b} $$

where Q0 is the total injected amount of tracer (mg), C1 and b are the intercept and the slope of the linear regression equation between plasma concentration and time (t), respectively.

The area under the curve (C1/b) calculated by this algorithm was underestimated; therefore, the calculated clearance value was corrected by the Brochner-Mortensen’s formula and standardized for a body-surface area (BSA) of 1.73 m2 (the Haycock’s formula) to derive the final GFR [7, 19].

Single-sample method (Cl ss)

The formula described by Jacobsson [8] was based on corrections for non-immediate mixing and non-uniform distribution of the tracer. The distribution volume was calculated as a function of the body weight.

$$ {Cl}_{ss}=\frac{1}{\mathrm{t}/\mathrm{V}+0.0016}\times \ln \left(\frac{{\mathrm{Q}}_0}{\mathrm{V}\times {\mathrm{C}}_{\mathrm{t}}}\right) $$

where Q0 is the total injected amount of tracer (mg), t is the time interval between injection and sampling (min), Ct is the iodine concentration in the plasma sample taken at the time (t), and V is the calculated distribution volume (ml) of the participant. The clearance values were adjusted to 1.73 m2 body surface.

The reference values of GFR were measured as the 2-, 3- and 4-h three-point plasma clearance of iohexol standardized to 1.73 m2 BSA (3 pt. iGFR), as described above, also denoted as mGFR (measured GFR). DPSM-GFR and SPSM-GFR were expressed as GFRix or GFRi, i and x impliy the sampling time (hr) of the blood used to calculate the corresponding plasma clearance of iohexol. For example, DPSM-GFR determined by the 2- and 3-h samples was denoted as GFR23, and SPSM-GFR determined by the single 2-h sample was denoted as GFR2, and so on.

Statistical analysis

Baseline characteristics are presented as the mean ± SD for continuous variables and as n (%) for categorical variables.

Taking 3 pt. iGFR as the reference standard (mGFR), bias was assessed as the median difference and precision was assessed as the interquartile range (IQR) for the difference. Accuracy was assessed as the percentage of estimates that differed within 30 and 10% of the mGFR (i.e. P30 and P10). Agreement between GFRix and mGFR was reported according to Lin’s concordance correlation coefficient (CCC) [20]. Moreover, the agreement between different methods was evaluated graphically by plotting the ratios of GFRix/mGFR against mGFR according to Bland and Altman [21].

Confidence intervals (CI) were calculated by means of bootstrap methods (1000 bootstraps). The significance of the differences among different methods was determined with the use of the signed-rank test for bias, the bootstrap method for the interquartile range from the 1000 bootstrap samples, and McNemar’s test for P30 and P10. P values < 0.05 were considered significant. Analyses were performed with the use of SPSS for Windows 21.0 (SPSS Inc., USA) and MedCalc version 18 (Medcalc software, Belgium).

Results

The detailed results of each different sampling method with individual correlations against mGFR are shown in Table 1. Overall, the best correlation from the analyses carried out was obtained from GFR24, the 2- and 4-h blood sampling, with a CCC of 0.9988, which indicates almost perfect agreement with the 3 pt. iGFR. Moreover, GFR24 showed the best accuracy: none of the estimates differed by more than 30% from the mGFR and only one (0.06%) estimate differed by more than 10% (P30, 100%; P10, 99.4%). Bias was very low and similar among DPSM-GFRs and SPSM-GFRs, while the precision of GFR24 was significantly better with an IQR of the difference of 1.52 ml/min/1.73m2.

Table 1 Summary of results of correlations between different sampling methods and three-point plasma clearance of iohexol

The agreement of each sampling method predicting mGFR is graphically illustrated in Figs. 1 and 2, with the solid lines and the dashed lines delineating the boundaries defined by P30 and P10, respectively. The percentage difference of GFR24 from the mGFR was fairly stable throughout the whole range of measured GFR. With respect to the other methods, the dots become increasingly scattered as GFR decreases, especially the SPSM-GFRs at a GFR level of < 60 ml/min/1.73m2.

Fig. 1
figure 1

Agreement between DPSM and three-point plasma clearance. Notes: The solid lines and the dashed lines delineate the boundaries defined by P30 and P10, respectively. GFRix: the corresponding DPSM-GFR determined by two samples drawn at the time i and x (hr) after injection. mGFR: the 2-, 3- and 4-hr three-point plasma clearance of iohexol. Units are in ml/min/1.73m2

Fig. 2
figure 2

Agreement between SPSM and three-point plasma clearance. Notes: The solid lines and the dashed lines delineate the boundaries defined by P30 and P10, respectively. GFRi: the corresponding SPSM-GFR determined by single sample drawn at the time i (hr) after injection. mGFR: the 2-, 3- and 4-hr three-point plasma clearance of iohexol. Units are in ml/min/1.73m2

Table 2 shows the performance of the DPSM-GFRs and SPSM-GFRs at different measured GFR levels. Generally, the correlation metrics were not significantly different between DPSM and SPSM when GFR was above 60 ml/min/1.73m2. However, SPSM showed noticeably imprecise and poor accuracy compared to DPSM in the GFR < 60 ml/min/1.73m2 subgroup.

Table 2 Performance of dual- and single- sampling methods at different GFR levelsa

Discussion

In this paper, we have presented a comparison of 2 commonly used plasma sampling methods for GFR measurements: single- and dual-plasma sampling method. The data demonstrated well that DPSM using blood samples drawn at 2- and 4-h post-injection obtained the best correlation in terms of both precision and accuracy.

These results were highly consistent with the early study done by Waller DG [15], who compared alternative methods including SPSM, DPSM and external detector clearance rate using 2–5 h samples. The 2- and 4-h blood sampling correlates excellently with multiple-point plasma clearance (r = 0.996) with a standard error of 2.8 ml/min/1.73m2. Similar results were found by Russell et al. [22], who compared the DPSM technique with a two-compartment GFR assessment.

In fact, the BNMS guidelines [23] recommend a slope-intercept method requiring between two and four samples in the conclusion that “the majority of literature suggests that the single-sample method is less precise than the slope-intercept technique.” On the other hand, the plasma clearance measurement recommended by the current international guidelines [24] is the single-sample technique for clinical measurement of GFR in patients with GFR ≥ 30 ml/min/1.73m2 based on the Groth 4-h methodology [10]. However, both of the guidelines were written decades ago.

It should be noted from Table 1 that all the studied methods had a low bias and yielded optimal agreement with the 3 pt. iGFR, while the performance in precision and accuracy varied substantially regarding a certain method or GFR level (Table 2). This is partially because positive errors and negative errors cancel each other out, and thus blunt the real deviation. This phenomenon can be even more obviously observed at a lower GFR level and SPSM using a too-early sample (2-h SPSM, see Fig. 2). The explanation for this is well documented in the literature [4, 25, 26]: a terminal monoexponential clearance is not reached even at 4 h and therefore some AUC is missing from the calculation, causing the GFR to be overestimated [27]. However, DPSM suffers from the same deficiency rooted in the mono-compartment model, which results in a tendency towards overestimation when GFR was low and the opposite when GFR was high [28].

Prolongation of the sampling time is the main strategy for avoiding this error, which is crucial in SPSM. Indeed, in our study, by using a later blood sampling (4 h), the precision of SPSM was greatly improved as the IQR of the difference decreased from 21.99 to 4.93 ml/min/1.73m2, while the accuracy within P30 increased from 55.2 to 82.8%. Conventional practice to choose proper sampling time is based on the expected GFR before testing. Jacobsson calculated the optimal sampling time for clearance values around 100 ml/min to be 3 h and for clearance values around 30 ml/min to be 10 h [8]. For severe renal dysfunction (expected GFR < 15 ml/min/1.73m2), a blood sample drawing at 24 h is suggested [24]. Recently, a systematic review [27] compared SPSM results with a gold standard nine-point AUC measurement of GFR as well as slope-intercept methods (including the 2-, 3- and 4-h three-point plasma clearance and the 2- and 3-h DPSM) for 412 published GFR studies and concluded that the method described by Fleming [14] is the best SPSM and provides equivalent accuracy and precision to the slope-intercept-GFR.

The limitation of this study is that we lack data beyond 4 h, which will result in slightly overestimation of the GFR, as discussed above. Schwartz et al. [4] examined the plasma disappearance curve of iohexol in 27 children to determine the degree of overestimation in GFR due to shortening sampling time from 6 to 5 and 4 h. According to the multi-point AUC measurement, the authors found a significant 3% overestimation if sampling time was truncated at 4 h post-injection. However, the differences did not become much larger when only lower GFR values were examined and a 3% overestimation is probably clinically irrelevant as discussed by the authors. Moreover, it was demonstrated in their work that the area under the slow curve of the 4 h study was not significantly different from 6 h, which is to say, the AUC used to calculate DPSM-GFR was not significantly different between 4- and 6-h.

Conclusions

In conclusion, clearance values based upon three-point plasma samples can be substituted by 2- and 4-h DPSM. When GFR was ≥60 ml/min/1.73m2, 4-h SPSM demonstrated approximate accuracy and thus can be recommended for clinical measurement of GFR. For patients with expected GFR < 60 ml/min/1.73m2, a prolonged sampling time is more reliable.

Abbreviations

3 pt iGFR:

The 2-, 3- and 4-h 3-point plasma clearance of iohexol

AUC:

Area under the curve

BNMS:

British Nuclear Medicine Society

BSA:

Body-surface area

CCC:

Concordance correlation coefficient

CI:

Confidence interval

CKD:

Chronic kidney disease

DPSM:

Dual-plasma sampling methods

GFR:

Glomerular filtration rate

HPLC:

High-performance liquid chromatography

IQR:

Interquartile range

mGFR:

Measured glomerular filtration rate

SPSM:

Single-plasma sampling methods

References

  1. Brown SC, O'Reilly PH. Iohexol clearance for the determination of glomerular filtration rate in clinical practice: evidence for a new gold standard. J Urol. 1991;146(3):675–9.

    Article  PubMed  CAS  Google Scholar 

  2. Stevens LA, Levey AS. Measured GFR as a confirmatory test for estimated GFR. J Am Soc Nephrol. 2009;20(11):2305–13.

    Article  PubMed  Google Scholar 

  3. Soveri I, Berg UB, Björk J, et al. Measuring GFR: a systematic review. Am J Kidney Dis. 2014;64(3):411–24.

    Article  PubMed  Google Scholar 

  4. Schwartz GJ, Abraham AG, Furth SL, Warady BA, Muñoz A. Optimizing iohexol plasma disappearance curves to measure the glomerular filtration rate in children with chronic kidney disease. Kidney Int. 2010;77(1):65–71.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  5. Frennby B, Sterner G, Almén T, Hagstam KE, Hultberg B, Jacobsson L. The use of iohexol clearance to determine GFR in patients with severe chronic renal failure--a comparison between different clearance techniques. Clin Nephrol. 1995;43(1):35–46.

    PubMed  CAS  Google Scholar 

  6. Sterner G, Frennby B, Hultberg B, Almen T. Iohexol clearance for GFR-determination in renal failure--single or multiple plasma sampling. Nephrol Dial Transplant. 1996;11(3):521–5.

    PubMed  CAS  Google Scholar 

  7. Bröchner-Mortensen J. A simple method for the determination of glomerular filtration rate. Scand J Clin Lab Invest. 1972;30(3):271–4.

    Article  PubMed  Google Scholar 

  8. Jacobsson L. A method for the calculation of renal clearance based on a single plasma sample. Clin Physiol. 1983;3(4):297–305.

    Article  PubMed  CAS  Google Scholar 

  9. Russell CD, Bischoff PG, Kontzen FN, et al. Measurement of glomerular filtration rate: single injection plasma clearance method without urine collection. J Nucl Med. 1985;26(11):1243–7.

    PubMed  CAS  Google Scholar 

  10. Christensen AB, Groth S. Determination of 99mTc-DTPA clearance by a single plasma sample method. Clin Physiol. 1986;6(6):579–88.

    Article  PubMed  CAS  Google Scholar 

  11. Watson WS. A simple method of estimating glomerular filtration rate. Eur J Nucl Med. 1992;19(9):827.

    Article  PubMed  CAS  Google Scholar 

  12. Itoh K, Tsushima S, Tsukamoto E, Tamaki N. Reappraisal of single-sample and gamma camera methods for determination of the glomerular filtration rate with 99mTc-DTPA. Ann Nucl Med. 2000;14(3):143–50.

    Article  PubMed  CAS  Google Scholar 

  13. Hackstein N, Langheinrich AC, Rau WS. Lopromide one-sample clearance as a measure of glomerular filtration rate. Clin Physiol Funct Imaging. 2002;22(2):99–107.

    Article  PubMed  CAS  Google Scholar 

  14. Fleming JS, Persaud L, Zivanovic MA. A general equation for estimating glomerular filtration rate from a single plasma sample. Nucl Med Commun. 2005;26(8):743–8.

    Article  PubMed  Google Scholar 

  15. Waller DG, Keast CM, Fleming JS, Ackery DM. Measurement of glomerular filtration rate with technetium-99m DTPA: comparison of plasma clearance techniques. J Nucl Med. 1987;28(3):372–7.

    PubMed  CAS  Google Scholar 

  16. Hansen HP, Rossing P, Mathiesen ER, Hommel E, Smidt UM, Parving HH. Assessment of glomerular filtration rate in diabetic nephropathy using the plasma clearance of 51Cr-EDTA. Scand J Clin Lab Invest. 1998;58(5):405–13.

    Article  PubMed  CAS  Google Scholar 

  17. Peters AM, Henderson BL, Lui D, Blunkett M, Cosgriff PS, Myers MJ. Appropriate corrections to glomerular filtration rate and volume of distribution based on the bolus injection and single-compartment technique. Physiol Meas. 1999;20(3):313–27.

    Article  PubMed  CAS  Google Scholar 

  18. Piepsz A, Ham R, De Sadeleer C. Guidelines for the measurement of glomerular filtration rate using plasma sampling. Nucl Med Commun. 2005;26(2):175–6. author reply 176-8

    Article  PubMed  Google Scholar 

  19. Haycock GB, Schwartz GJ, Wisotsky DH. Geometric method for measuring body surface area: a height-weight formula validated in infants, children, and adults. J Pediatr. 1978;93(1):62–6.

    Article  PubMed  CAS  Google Scholar 

  20. Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45:255–68.

    Article  PubMed  CAS  Google Scholar 

  21. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.

    Article  PubMed  CAS  Google Scholar 

  22. Russell CD. Optimum sample times for single-injection, multisample renal clearance methods. J Nucl Med. 1993;34(10):1761–5.

    PubMed  CAS  Google Scholar 

  23. Fleming JS, Zivanovic MA, Blake GM, Burniston M, Cosgriff PS. Guidelines for the measurement of glomerular filtration rate using plasma sampling. Nucl Med Commun. 2004;25(8):759–69.

    Article  PubMed  Google Scholar 

  24. Blaufox MD, Aurell M, Bubeck B, et al. Report of the radionuclides in Nephrourology committee on renal clearance. J Nucl Med. 1996;37(11):1883–90.

    PubMed  CAS  Google Scholar 

  25. Bröchner-Mortensen J, Rödbro P. Optimum time of blood sampling for determination of glomerular filtration rate by single-injection [51Cr]EDTA plasma clearance. Scand J Clin Lab Invest. 1976;36(8):795–800.

    Article  PubMed  Google Scholar 

  26. Moore AE, Park-Holohan SJ, Blake GM, Fogelman I. Conventional measurements of GFR using 51Cr-EDTA overestimate true renal clearance by 10 percent. Eur J Nucl Med Mol Imaging. 2003;30(1):4–8.

    Article  PubMed  CAS  Google Scholar 

  27. McMeekin H, Wickham F, Barnfield M, Burniston M. A systematic review of single-sample glomerular filtration rate measurement techniques and demonstration of equal accuracy to slope-intercept methods. Nucl Med Commun. 2016;37(7):743–55.

    Article  PubMed  CAS  Google Scholar 

  28. Thomsen HS, Vestergaard A, Golman K, Dorph S. One or two samples for determination of total plasma clearance of a nonionic contrast medium in patients undergoing enhanced CT. Acta Radiol. 1992;33(6):588–91.

    Article  PubMed  CAS  Google Scholar 

Download references

Availability of data and materials

The datasets generated and/or analyzed during the current study are available upon request from the corresponding author.

Author information

Authors and Affiliations

Authors

Contributions

Research idea and study design: ZS and LZ; blood sampling and HPLC essay: YZ, ZY, TFL and WYF; data acquisition/statistical analysis: YZ and ZS; supervision or mentorship: WYF and LZ. YZ is a major contributor in writing the manuscript, and ZS critically revised the manuscript and added some important intellectual content, the two authors contributed equally to this work. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Li Zuo.

Ethics declarations

Ethics approval and consent to participate

This is a cross-sectional study conducted at the Peking University People’s Hospital (PUPH) from Oct 2015 to Feb 2017 and approved by the PUPH ethics committee (2014PHB098–01). All volunteers were informed and signed the consent form.

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, Y., Sui, Z., Yu, Z. et al. Accuracy of iohexol plasma clearance for GFR-determination: a comparison between single and dual sampling. BMC Nephrol 19, 174 (2018). https://doi.org/10.1186/s12882-018-0965-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12882-018-0965-7

Keywords